Mostrar el registro sencillo del ítem
| dc.contributor.author | Guale Coello, Angie Arlette | |
| dc.contributor.author | Rendón Morán, Michael Guillermo, Director | |
| dc.date.accessioned | 2026-07-02T16:16:26Z | |
| dc.date.available | 2026-07-02T16:16:26Z | |
| dc.date.issued | 2026 | |
| dc.identifier.citation | Guale Coello A.A. (2026) Desarrollo y validación de un método de análisis por UPLC-DAD para la determinación de ERITROSINA (E127) en bebidas de consumo humano [Tesis Maestría] Escuela Superior Politécnica del Litoral. Guayaquil, 108 páginas | es_EC |
| dc.identifier.uri | http://www.dspace.espol.edu.ec/handle/123456789/69251 | |
| dc.description | El presente trabajo tuvo por objetivo determinar la conformidad de un método UPLC-DAD para la cuantificación de eritrosina (E127) en bebidas gaseosas y rehidratantes frente a los requisitos de validación establecidos en la norma ISO/IEC 17025:2017. Para ello, se desarrolló un procedimiento cromatográfico orientado a la identificación y cuantificación del analito, empleando curvas de calibración en el rango de 0,05–2,50 μg/mL y un esquema experimental aplicado y controlado. La validación se efectuó siguiendo lineamientos internacionales (Eurachem e ISO/IEC 17025) y consideró selectividad, linealidad, límites de detección y cuantificación, precisión (repetibilidad y precisión intermedia), exactitud, robustez e incertidumbre de medida. El método mostró un comportamiento lineal consistente (R² entre 0,9970 y 0,9995), ausencia de interferencias relevantes en el tiempo de retención del analito y un desempeño adecuado de precisión y exactitud para su uso rutinario. Los LOD/LOQ fueron 0,0193/0,0642 μg/mL en rehidratante y 0,0117/0,0391 μg/mL en gaseosa, con incertidumbre expandida relativa de 7,97% (k=2). Finalmente, se aplicó el método a 12 bebidas del mercado ecuatoriano, obteniéndose resultados no detectados (<LOD), lo que confirma su aptitud para control analítico. | es_EC |
| dc.description.abstract | The present study aimed to determine the conformity of a UPLC–DAD method for the quantification of erythrosine (E127) in carbonated and rehydration beverages with the validation requirements established in ISO/IEC 17025:2017. For this purpose, a chromatographic procedure was developed for the identification and quantification of the analyte, using calibration curves in the range of 0.05–2.50 μg/mL and a controlled experimental design. Method validation was performed following international guidelines (Eurachem and ISO/IEC 17025:2017) and included selectivity, linearity, limits of detection and quantification, precision (repeatability and intermediate precision), accuracy, robustness, and measurement uncertainty. The method showed consistent linear behavior (R² between 0.9970 and 0.9995), absence of relevant interferences at the analyte retention time, and adequate precision and accuracy for routine use. The LOD/LOQ values were 0.0193/0.0642 μg/mL in rehydration beverages and 0.0117/0.0391 μg/mL in carbonated beverages, with a relative expanded uncertainty of 7.97% (k = 2). Finally, the method was applied to the analysis of 12 beverages from the Ecuadorian market, yielding non-detected results (<LOD), which confirms its suitability for analytical control. Keywords: Erythrosine (E127), UPLC–DAD, analytical validation, ISO/IEC 17025, beverages. | es_EC |
| dc.publisher | ESPOL.FCNM | es_EC |
| dc.subject | Eritrosina (E127) | es_EC |
| dc.subject | UPLC-DAD | es_EC |
| dc.subject | Validación analítica | es_EC |
| dc.subject | ISO/IEC 17025 | es_EC |
| dc.subject | Bebidas | es_EC |
| dc.title | Desarrollo y validación de un método de análisis por UPLC-DAD para la determinación de ERITROSINA (E127) en bebidas de consumo humano | es_EC |
| dc.type | Thesis | es_EC |
| dc.identifier.codigoespol | T-116338 | |
| dc.identifier.codigo | POSTG067 |